Background: Why Feed Trace Element Testing Matters

Meeting the specification for copper, iron, zinc and manganese is a mandatory entry condition for feed producers applying for a premix production licence. These four elements are essential trace nutrients for animals: adequate and accurate levels support normal physiological metabolism, improve production performance, strengthen immunity and maintain animal health.

Reliable testing also satisfies regulatory compliance and controls several distinct risks at once. It prevents trace element overdose and the animal toxicity that follows, reduces the soil and water pollution caused by excess excretion through manure - in line with the green, low-carbon direction of the feed industry - and controls elemental residues in livestock products so they cannot migrate up the food chain to humans. In short, accurate trace element data protects both animal health and public food safety.

Project Objective

A feed quality laboratory needed to determine copper, iron, zinc and manganese in compound feed samples by flame atomic absorption spectrometry. The laboratory required a system that could deliver stable atomization, a wide linear working range for each element, low short-term drift and repeatable results across a production testing workflow. The work was performed on a Wanyi AA2310 atomic absorption spectrophotometer running WYatom software.

Instrumentation

No.Item
1Wanyi AA2310 atomic absorption spectrophotometer
2Air compressor
3WYatom workstation software
4Copper hollow cathode lamp
5Iron hollow cathode lamp
6Zinc hollow cathode lamp
7Manganese hollow cathode lamp

Single element stock standards of copper, iron, zinc and manganese at 1000 mg/L were used to prepare the calibration series.

Calibration and Linearity

Each element was calibrated with a blank plus five concentration levels, and every level was injected in triplicate to monitor short-term stability. All four calibration curves were fitted with a linear model and intercepts were calculated rather than forced through zero.

ElementRange (mg/L)Curve equationCorrelation R
Cu0.2 - 1.0y = 0.270424x + 0.0046830.9999
Fe0.5 - 4.0y = 0.123631x + 0.0195230.9996
Zn0.2 - 1.0y = 0.386340x + 0.0209950.9990
Mn0.1 - 0.5y = 0.494343x + 0.0059480.9997

Absorbance increased proportionally with concentration across every range, and the correlation coefficients of all four elements were 0.999 or better, confirming the linearity and quantitative reliability required for routine feed testing.

Typical Absorbance Response

ElementBlankL1L2L3L4L5
Cu (Abs)0.00680.05760.11380.16740.22200.2739
Fe (Abs)0.00460.07430.14960.27040.38850.5129
Zn (Abs)0.00010.09160.18080.25740.33180.4024
Mn (Abs)0.02440.05450.10380.15730.20390.2516

Repeatability

Three replicate injections were made at every calibration level. RSD for all four elements remained below 0.2 percent, with most points inside 0.15 percent, demonstrating very stable flame atomization and a robust autosampler-free manual workflow.

Real Sample Analysis

A compound feed sample (sample 2) was analysed for all four elements. A 2.0076 g test portion was digested and made up to 50 mL, with element-specific dilution factors applied before measurement. All results fell within the expected range for supplemented compound feed.

ElementTest portion (g)Volume (mL)DilutionMeasured (mg/L)Content (mg/kg)
Cu2.00765010.890322.2
Fe2.00765022.8192140.0
Zn2.00765050.774896.5
Mn2.00765050.228228.4

Copper was quantified directly without dilution, while iron was measured at a two-fold dilution and zinc and manganese at five-fold dilution to bring them into the calibrated range - a practical demonstration of how wide this flame AAS linear window is for feed matrices.

Conclusion

The correlation coefficients of copper, iron, zinc and manganese were all 0.999 or above, and three-replicate RSD was below 0.2 percent. The method is accurate, reliable and sensitive, and the Wanyi AA2310 flame atomic absorption system is fully capable of determining copper, iron, zinc and manganese in feed samples. It gives feed producers a dependable analytical basis for premix licence applications, raw material and finished product screening, and routine quality control.

Note: hydrochloric acid used in sample preparation is a strong acid and must be handled in a fume hood with appropriate personal protective equipment.